•  15inch Touch Screen Medical Ventilator Covid19 hospital respirator
15inch Touch Screen Medical Ventilator Covid19 hospital respirator

Product Introduction

Item No:  JQ-1100A
Ventilation mode                           
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV,SIGH,MANUAL

Ventilator parameter 
Tidal volume (Vt) 0 ~ 2000 ML
Frequency (Freq) 1 /min ~ 100 /min
Oxygen concentration 21 % ~ 100 %
I:E 4 : 1 ~ 1 : 8
PEEP 0 cmH2O ~ 40 cmH2O
Pressure triggering sensitivity (Ptr)  -20 cmH2O ~ 20 cmH2O (Based on PEEP)
Flow trigger sensitivity (Ftr) 0.5 L/min ~ 30 L/min
Pressure control (PC) 5 cmH2O ~ 80 cmH2O  
Pressure support (PS) 0 cmH2O ~ 80 cmH2O  
SIGH 0 (off) 1/100 ~ 5/100
Apnea ventilation OFF, 5 s ~ 60 s
Pressure limit 20 cmH2O ~ 100 cmH2O
Monitoring parameter 
Frequence(Freq) 0 /min ~ 100 /min 
Tidal volume (Vt) 0 mL ~ 2500 mL 
MV 0 L/min ~ 99 L/min
Airway pressure 0 cmH2O ~ 100 cmH2O
Dynamic lung compliance monitoring  1 mL/cmH2O ~ 1000 mL/cmH2O  
ETCO2 Concentration 0 mmHg ~ 152 mmHg (0%-20%)  
Oxygen concentration 15 % ~ 100 %
Oscillogram
P-T(Pressure-Time)
F-T(Flow-Time)
V-T(Volume-Time
ETCO2 -T (End-Tidal CO2 -Time)
P-V Loop(Pressure-Volume Loop)
F-V Loop(Flow-Volume Loop)
P-F Loop(Pressure-Flow Loop)
Alarm and protection 
The AC power failure alarm Power failure or no connection  
Internal battery backup low voltage alarm  < 11.3 ± 0.3 V
No tidal volume No tidal volume within 6 s
High minute volume alarm 5 L/min ~ 99 L/min
Low minute volume alarm 1 L/min ~ 30 L/min
High airway  pressure alarm 20cmH2O ~ 100cmH2O  
Low airway  pressure alarm 0cmH2O ~ 20cmH2O
High oxygen concentration alarm 19 % ~ 100 %
Low oxygen concentration alarm 18 % ~ 99 %
Continuous pressure alarm (PEEP + 1.5 cmH2O) over 16s  
Suffocation warning 5 ~ 60 s
Fan error Show on screen
Oxygen deficit Show on screen
The maximum limited pressure  <12.5 kPa
Working condition
Gas source O2,  AIR
Pressure 280 kPa ~ 600 kPa
Voltage 220 V ± 22 V
Power frequency 50 Hz ± 1 Hz
Input power 900 VA (With air compressor)
  250 VA (Without air compressor